Measuring current, max. | 1 – 2000 A |
Pulse energy, max. | 1750 J |
Pulse width, max. | 3 µs – 1 s adjustable |
Pulse voltage | 10 – 400 V adjustable |
Resistance measurement | 0.01 mΩ – 35 Ω |
Dimensions | 365 (W) x 325 (L) x 167 (H) mm³ |
The inductance curve of test specimens with an ohmic resistance that exceeds the measuring range can nevertheless be measured when the resistance is entered manually.
For very large power inductors, that store a larger amount of energy, the model with the option /E with double pulse energy capability is available.
Description and technical specifications of the Power Choke Tester DPG10
Scope of delivery:
- DPG10 – 2000B/E
- Cable set
- Test leads 2x Force, 1x Sense, 0.6 m each with alligator clips
- USB cable and RS232 cable
- Power cord
- PC software for Windows
- Instruction manual
- Calibration certificate
- Library (DLL) for integration of the instrument in automated test environments using your own application or LabVIEW
- Kelvin test leads (KK11-4)
- Kelvin test leads for small devices under test (KK12-4)
- Kelvin test leads up to 5 kA (KK11-6)
- Extra long test leads (KL11-4)
- Extra long test leads (KL11-6)
- Hard-top case (CASE1)
- Customer specific test leads on request